56
Operation Procedure
Setting VTBI
Set the VTBI for the solution delivery (Default: disabled).
To activate this feature, contact TERUMO trained service technicians.
Setting VTBI to enable (ON) also sets VTBI time to enabled (ON).
Display
select switch
1)
Press the Display select switch to select the VTBI display.
Notes
• The VTBI tab
is selected.
• The flow rate display shows the VTBI display and the VTBI time display.
• The VTBI display appears in dark blue (selected).
Flow rate tab
VTBI tab
VTBI display
VTBI time display
Increases value
Dial
Decreases
value
2)
Use the Dial to adjust the VTBI.
Note
• When the flow rate has already been entered, the VTBI time display indicates the
remaining time and VTBI time calculated from the flow rate and VTBI.
Start up icon
VTBI time
Remaining time
Notes
• The VTBI can be set from 0.10 to 9999.00 and “----.--” (Unlimited).
• To set to unlimited, turn one rotation or more in the direction of decreasing the Dial when VTBI is “0”.
• In the case of a solution delivery with no setting for the VTBI (“----.--”: Unlimited), the pump will not stop until the Syringe Empty
is issued due to running out drug solution, and therefore the product should be stopped before running out of the drug solution.
• The VTBI can also be set in mass units. In this case, VTBI can be set from 0.01 ng to 9999.99 kg and “----.--” (Unlimited). (When
the selecting dose mode (weight mode/drug library) is set.)
• If not operated for 10 seconds, the input area moves to the flow rate display.
• VTBI can be changed only when delivery is stopped.
• Rotating the Dial while pressing the Stop switch allows changing the setting at the rate of 100 times.