15
Parts Description
Common screen
Menu screen
For the programming procedure, see page 79.
Menu 1
Menu 2
Menu tab
Menu details display
Menu tab
Menu details display
Menu
Menu tab
Menu details
Note
Menu 1
Press and hold the Display select
switch for 2 seconds or more.
This moves the input area to the
next item before displaying the menu
screen.
Display
select switch
2 seconds
or more
Dose mode select
mL/h
μ
g/kg/min
mg/kg/h
mg/kg/h+DIPRIVAN
Library
TCI
Other dose units can be
specified in the Drug Library.
“Library” is displayed after
uploading the Drug Library.
Occlusion
detection pressure
1 to 10
(± 10 kPa
↔
± 120 kPa)
Sets the occlusion detection
pressure level.
Keypad lock
ON
OFF
Sets the keypad lock function
to ON or OFF.
Bolus
Hands On Bolus
Hands Free Bolus
Bolus infusion is available with
the Purge switch.
TCI
(TCI mode only)
Maximum flow rate
Plasma limit
Sets the upper limit of flow rate
and plasma concentration.
Advanced dose
mode
OFF
Interval/Intermittent Mode
Multi-step Mode
Delayed Start Mode
Selects an advanced dose
mode. (Displayed when a
drug library that allows for
advanced dose mode is
selected.)
Combination dose
mode
―
For details, contact TERUMO
trained service technicians.