124
About External Communication Functions
Item
Size (Byte)
Details
Power status
1
b0-b3 (low order)
Battery level
0x0: (Reserved)
0x1:
Remaining battery level, 1 bar
0x2:
Remaining battery level, 2 bars
0x3:
Remaining battery level, 3 bars
0x4:
Remaining battery level, 4 bars
0x5:
Remaining battery level, 5 bars
0x6-0xF: (Reserved)
b4-b5
Power type
0x0: Battery
drive
0x1: AC
drive
0x2-0x3: (Reserved)
b6-b7 (high order)
Sub-battery status
0x0: (Reserved)
0x1: Sub-battery,
normal
0x2: Sub-battery,
abnormal
0x3: (Reserved)
Operation monitor
1
b0-b1 (low order)
Occlusion monitor (0x0: 0%, 0x1: 25%, 0x2: 50%,
0x3: 75%)
b2
(Reserved)
b3
Alarm sound status (0: Without sound, 1: With
sound)
b4
Keypad lock status (0: Not locked, 1: Locked)
b5
Plunger status monitor (0: Set, 1: Not set)
b6
(Reserved)
b7 (high order)
(Reserved)
One bit used for b2-b7. (“1” indicates active.)
Syringe status
1
0x0
None
0x1
(Reserved)
0x2
5 mL
0x3
10 mL
0x4
20 mL
0x5
30 mL
0x6
50/60 mL
0x7-0xFF
(Reserved)
Flow rate
3
Value of 100 times the set flow rate
Volume delivered
3
Value of 100 times the volume delivered
VTBI
3
Value of 10 times the VTBI
Infinite: 0xFFFFFF
Invalid value (when the function is disabled): 0xEEEEEE
Dosage rate
3
Value of 100 times the dosage
Unit of dosage rate
1
0x00: mL/h
0x04:
μ
g/kg/min
0x08: mg/kg/h
Weight
2
Value of 10 times the weight
Invalid value (when the function is disabled): 0xFFFF
Dilution
3
Value of 100 times the dilution
Invalid value (when the function is disabled): 0x000000