41
Operation Procedure
Changing the target concentration (Cpt or Cet)
The target concentration (Cpt or Cet) can be changed during solution delivery.
Increases value
Dial
Decreases
value
1)
During solution delivery, turn the Dial with the graph or text
screen displayed to change the target concentration (Cpt or
Cet).
Note
• The confirmation message “Press Start to Confirm” is displayed on the LCD.
Start switch
Operation indicator (green flashing)
2)
Press the Start switch.
Notes
• The target concentration (Cpt or Cet) is confirmed.
• The buzzer sounds.
• The operation indicator flashes green.
• The line indicating the target concentration (Cpt or Cet) in the graph is updated.
Caution
• If the Start switch is not pressed within approximately 5 seconds, the setting is not applied and the previously set value is
restored. Be sure to press the Start switch to apply the setting in order to achieve the expected performance.