Changing Settings on the Menu
Advanced Dose Mode
An advanced dose mode can be selected from the three advanced dose modes in the table below according to the selected drug.
A library should be registered before selecting the advanced mode. Use the dedicated editing software (TERUFUSION Drug
Library Manager or TERUFUSION Software Package) for registration. See page 66 for how to select the library.
Advanced dose mode
Details
Interval/Intermittent
Mode
Two specified dosage rates are alternately delivered at regular intervals for the specified number of steps.
. . . . .
5) Number of steps
Flow rate
2) Dosage rate 2
1) Dosage rate 1
3) 1 cycle time
4)
4) Time interval
Dosing time
Input parameters
1) Dosage rate 1
For the dosage rate range, see page 68.
When using an advanced dose mode, 0.00 can
be set for the dosage rate.
2) Dosage rate 2 (Displays: Final Flow Rate)
3) 1 cycle time (Displays: Time/Cycle)
00h01min to 23h59min
4) Time interval (Displays: Time)
00h01min to 23h59min
5) Number of steps (Displays: Number of Steps)
1 to 255 steps
Multi-step Mode
The speed of solution delivery increases or decreases at the specified levels (up to 10 levels).
. . . . .
Flow rate
2) Dosage rate 2
1) Dosage rate 1
4) Number of steps
Dosing time
3) Time
Input parameters
1) Dosage rate 1
For the dosage rate range, see page 68.
When using an advanced dose mode, 0.00 can
be set for the dosage rate.
2) Dosage rate 2 (Displays: Final Flow Rate)
3) Time (Displays: Time to Final Flow Rate)
00h01min to 23h59min
4) Number of steps (Displays: Steps)
1 to 10 steps
Delayed Start Mode
Solution delivery starts after the specified time has elapsed.
Flow rate
2) Dosage rate 2
1) Dosage rate 1
3) Delay Time
Dosing time
Input parameters
1) Dosage rate 1
For the dosage rate range, see page 68.
When using an advanced dose mode, 0.00 can
be set for the dosage rate.
2) Dosage rate 2 (Displays: Final Flow Rate)
3) Delay Time (Displays: Delay Time)
00h01min to 23h59min
• “Dosage rate 1” can be set in the Dose Rate in the flow rate setting screen.
• There is no rule stating which of the volumes between dosage rate 1 and dosage rate 2 should be larger.
• If one of the dosage rate 1 or 2 is not set, solution delivery does not start.
• The dosage rate must be set within the setting range for each library.
• The calculation result of flow rate (mL/h) must be set within the flow rate setting range.
Note
• When registering library, determine the range of the dosage rate considering the drug and the dose method. (Depending on the
setting of the dosage rate, the occlusion may occur due to the thrombus formation within the infusion line.)