129
After Use
How to Clean Components
After use, clean all components as described below.
Caution
• After cleaning, completely dry the
AC inlet before use. Otherwise an
electric shock or short circuit may
occur resulting in failure of the
product.
Cleaning the AC inlet
Lightly wipe off dirt with a cotton swab
or damp cloth (not wet).
Cleaning the panel surface
Wipe with a damp (not wet) soft cloth.
Cleaning the outside of the syringe pump and the Dial
Wipe with a damp (not wet) soft cloth.
When disinfecting, use a designated disinfectant.
The Dial is fixed to the body of the pump with a
magnet. Remove it following the procedure below
before cleaning.
How to remove the Dial
1 Turn the groove of the Dial to the rear of the device.
2 Place a flathead screwdriver in the groove of the
Dial and push it outwards.
3 Remove the Dial and wipe clean the body and the
Dial.
Grease is applied to the shaft. Do not wipe it off.
4 When you have finished cleaning, fit the Dial back
onto the body.
5 After fitting, turn on the power.
Check that setting for flow rate etc., can be
performed.
Cautions
• Do not rub the panel surface with a hard object.
• Do not rub with a hard object such as forceps or with a sharp object etc.