126
About External Communication Functions
Item
Size (Byte)
Details
Plasma limit
1
200-350%, subtract 100 from the value (1% step)
Range: 0x64-0xFA
The upper limit undefined: 0x00
When the dose mode other than TCI is selected: 0x00
Maximum flow rate
3
Value of 100 times the maximum flow rate
0.00 to 999.00 (0.01 step)
Range: 0x000000-0x01863C
When the maximum flow rate is not set or when the dose mode other than
TCI is selected: 0x000000
Predicted plasma
concentration
3
Value of 100 times the concentration 0.00 to 50.00 (0.10 step)
Range: 0x000000-0x001388
When the dose mode other than TCI is selected: 0x000000
Predicted effect-site
concentration
2
Value of 100 times the concentration 0.00 to 50.00 (0.10 step)
Range: 0x0000-0x1388
When the dose mode other than TCI is selected: 0x0000
Predicted awakening
time
3
Hour:minute:second 00:00:00 to 99:59:59
Range: 0x000000-0x995959
When the initial flow rate is not set or when the dose mode other than TCI is
selected: 0x000000
Elapsed time
3
Hour:minute:second 00:00:00 to 99:59:59
Range: 0x000000-0x995959
Before the first solution delivery is started in the TCI mode or when the dose
mode other than TCI is selected: 0x000000
(Reserved)
2
(Reserved)
(Reserved)
2
(Reserved)
Volume delivered by
TCI
2
Mass values (mg or
μ
g) of volume delivered: 0 to 65535 (1 step)
Range: 0x0000-0xFFFF
When the dose mode other than TCI is selected: 0x0000
Volume unit delivered
by TCI
1
0x12:
μ
g, 0x13: mg
When the dose mode other than TCI is selected: 0x00
Total number of bytes
82
* For more details, contact TERUMO trained service technicians.