68
Other Operation Procedures
Confirmation switch
Increases value
Dial
Decreases
value
7)
If necessary, turn the Dial to set each parameter (amount,
solution, and weight), and press the Confirmation switch.
Notes
• When the dilution is set in the drug library, the dilution parameter cannot be
changed.
• Amount, solution and weight can be set within the following range.
Amount: 0.01 to 999.00*
Solution: 0.01 to 999.00*
Weight: 0.1 to 300.0 kg
* Can be set within the range of the dilution = Amount : 0.01 to 999.00
Solution
• With every press of the Confirmation switch, the input area moves to the next one.
• When the Back/Mute switch is pressed, the parameter input is cancelled and the
screen returns to the previous status.
• When all the parameters are input, the parameter confirmation screen appears.
• The background colour of the changed parameter value changes on the parameter
confirmation screen.
• When the Back/Mute switch is pressed on the parameter confirmation screen, the
parameter input is cancelled and the screen returns to that of step 5.
• Information of the patient and drugs can be checked, or parameter values can be
changed from the parameter confirmation screen.
Patient information screen
Drug information screen
Parameter value setting screen
(Amount/Solution)
Parameter value setting
screen (Weight)