67
Other Operation Procedures
Confirmation
switch
4)
Press the Confirmation switch.
Notes
• The screen changes to the library selection display.
• The selected library is displayed in dark blue.
Drug Library Version
Display select switch
Dial
5)
Turn the Dial or press the Display select switch to select the
library name, shown as "Drug2" in the example.
Note
• Turning the Dial or pressing the Display select switch moves the white frame.
Confirmation
switch
6)
Press the Confirmation switch.
When clinical advisory
*1
is set
Clinical advisory display
Notes
• If clinical advisory is set, a message appears. With every press of the Confirmation switch, the screen changes to the input
screen of amount and solution, or the input screen of dilution
*2
. If you want to set by Dilution, contact TERUMO trained service
technicians.
*
1
: Clinical advisory is a function to display a cautionary message about the selected library. The message can be registered
up to 120 characters. See the dedicated editing software (TERUFUSION Drug Library Manager or TERUFUSION Software
Package) for how to set the clinical advisory for each drug library.
*
2
: When the dilution is set in the drug library, the input screen of dilution appears.
• When the dose unit is mL/h, the screen changes to the parameter confirmation screen of step 8.