61
Operation Procedure
Stopping Solution Delivery
Stop switch
Operation indicator
1)
Press the Stop switch.
Notes
• The buzzer sounds and the operation indicator turns off, then the operation stops.
• The volume delivered stops increasing.
OFF
Close the three
way stopcock.
2)
Close the infusion line.
1
2
Clamp
3)
Pull the clamp and then turn.
Clutch
4)
While holding the clutch, remove the syringe.
Warning
• When removing the syringe installed in this product, first close the three way stopcock etc. of the infusion line. Otherwise
excessive infusion (siphoning free flow by gravity) of the drug solution may occur.