64
Clearing the Volume Delivered (except the TCI mode)
The volume delivered can be cleared.
Display
select switch
1)
Press the Display select switch repeatedly until the volume
delivered display is selected.
Note
• The volume delivered display appears in dark blue (selected).
Volume delivered display
Note
• The volume delivered can be cleared when solution delivery is in progress or when
stopped.
Confirmation
switch
2)
Press the Confirmation switch.
Note
• The buzzer sounds, the volume delivered is cleared to “0”, and the screen returns
to the flow rate screen.
Volume delivered
Notes
• When resuming the solution delivery after the value of the volume delivered is cleared to “0” mL, check the flow rate and the
VTBI, and re-set as required.
• To check the volume delivered after clearing the value, use the history function. (See page 100.)
• Turning the power off clears the volume delivered to “0”.
• If not operated for 10 seconds, the input area moves to the flow rate display.