66
Other Operation Procedures
Selecting Dose Mode (Drug Library)
The dose mode can be selected from the drug library.
Use the dedicated editing software (TERUFUSION Drug Library Manager or TERUFUSION Software Package) for registration. For
details, contact TERUMO trained service technicians.
Display
select switch
2 seconds
or more
1)
Press and hold the Display select switch for 2 seconds or
more when solution delivery is stopped.
Notes
• The menu 1 screen is displayed and then the dose mode select (Dose) display
appears.
• The set profile names are displayed next to “Library”.
• Profiles allow drugs to be classified according to their use and registered in the
pump. Selecting a profile can limit the available libraries. To change the profile, see
page 112.
Check
• A correct profile name is displayed.
Dose mode select tab
Profile name
Confirmation
switch
2)
Press the Confirmation switch.
Note
• The existing dose mode (light blue) changes to the selected status (dark blue).
Display select switch
Dial
3)
Turn the Dial or press the Display select switch to select the
library.
Note
• Turning the Dial or pressing the Display select switch moves the white frame.