34
Operation Procedure
Confirmation
switch
7)
Press the Confirmation switch.
Check
• The syringe brand and the syringe size are displayed on the LCD correctly.
Tube holder
Tube guide
8)
Insert the tube completely into the tube guide and set the
tube to the tube holder.
Warning
• When setting a syringe, ensure that the syringe plunger has been firmly set onto the slider hook and that the syringe has
been correctly set and that the flange of the syringe has been inserted into the flange holder. The elevation difference
between this product and the patient should be kept to a minimum. Drug solution may not be delivered correctly due to
rapid infusion caused by displacement of the syringe plunger from the slider hook, incorrect detection of the syringe size,
etc.
Cautions
• When setting a plunger into the slider, check that there is no gap between the plunger of the syringe and the slider. If the
plunger is set with gap between them, air or drug solution may be drawn in by the slider hook.
• If the LCD displays the syringe displacement icon and the buzzer sounds, check the syringe has been set correctly, and
re-set if necessary. If the syringe is not installed at the correct position, the solution delivery cannot be started.
• Even after the syringe is set in the proper position, if any fault is observed, immediately stop using the product and contact
TERUMO trained service technicians.
• When installing a syringe, make sure that the syringe size and brand displayed on the LCD correspond to the syringe. If
they do not correspond with each other, it may cause unintended solution delivery.
• When installing a syringe, pull forward the clamp and turn. Then, after the syringe is set, turn the clamp back to the
original position and gently push it to fix the syringe. In addition, the slider should be moved while holding the clutch. Any
unreasonable operation or excess operation will cause failure of this product.
• When using a syringe (specified sterile syringe for general use) for this product, ensure that the nominal capacity of the
syringe is not exceeded. If exceeded, the accuracy of flow rate or the alarm function cannot be guaranteed.