37
Operation Procedure
Increases value
Dial
Decreases
value
7)
Turn the Dial to adjust the target concentration (Cpt or Cet).
Caution
• When performing effect-site TCI, it is recommended to set the optimal plasma
concentration overshoot before starting. There may be the cases where
excessive overshoot in plasma concentration adversely affects the patient, in
particular, elderly patients or patients with high blood pressure.
Checks
The syringe pump is ready to start when the following are confirmed:
• The initial infusion dose and time are displayed.
• The Start Reminder message appears flashing in the lower section of the LCD.
• The buzzer sounds every minute of no operation. After the first solution delivery
starts, the buzzer sounds every 30 seconds of no operation.
• When the target concentration (Cpt or Cet) is 0.0, no message is displayed and no
buzzer sounds.
Initial infusion dose
Target concentration (Cpt or Cet)
Initial infusion time
Prediction lines
Notes
• The target concentration (Cpt or Cet) can be set within the following range:
0.0 to 15.0
μ
g/mL in 0.1
μ
g/mL step (Propofol)
0.0 to 50.0 ng/mL in 0.1 ng/mL step (Remifentanil)
• The default setting of the target concentration (Cpt or Cet) is 4.0
μ
g/mL (Propofol)
and 3.0 ng/mL (Remifentanil).
• If the set value exceeds 10.0
μ
g/mL (Propofol) or 10.0 ng/mL (Remifentanil), the
message “Soft upper limit reached, Override?” is displayed.
• When 15.0
μ
g/mL (Propofol) or 50.0 ng/mL (Remifentanil) is reached, the message
“This is the hard upper limit” is displayed on the LCD.
• As the value of target concentration (Cpt or Cet) is changed, the initial infusion
dose/time and the prediction lines in the graph change accordingly.