Setups
Reference
3Ć52
NOTE
When the compliance, explained in the next step, is set larger than
the measure range setting, the compliance is shifted to the meaĆ
sure range limit so that no more acquisition can be performed
when the current or voltage reaches the limit of measure range
setting. When a compliance is set to smaller that the measure
range, the measurement can only be performed within the limit of
compliance.
For measurement ranges, refer to Table 3Ć1 through Table 3Ć4.
Figure 3Ć21:ăVoltage and Current Range PopĆup Menu
ăStep 13:ă
Set compliance (see Figure 3Ć20).
Compliance is the upper or lower limit of the voltage or current that the
HPU or LPU can output, and is used to protect the DUT from too much
current or too much voltage. The voltage limit or current limit must be
set according to the I" mode or V" mode set for the SMU. Compliance
can be set using the [COMPLIANCE] selector, and must be set within
the ranges shown in Table 3Ć19 and 3Ć20.
The compliance set with the [COMPLIANCE] selector effects the meaĆ
sure range. Refer to step 12 for more information about the relation
between the compliance and measure range.
Summary of Contents for 372
Page 4: ......
Page 6: ...About this Manual ii ...
Page 14: ...Contents x ...
Page 22: ...Contents xviii ...
Page 30: ...Consignes de Sécurité Safety Summary xxvi ...
Page 31: ...Getting Started ...
Page 32: ......
Page 36: ...Overview Getting Started 1Ć4 ...
Page 90: ...Tutorial Ć About the Sample Disk Getting Started 1Ć58 ...
Page 91: ...Operating Basics ...
Page 92: ......
Page 108: ...Mounting the Device Under Test Operating Basics 2Ć16 ...
Page 110: ...Before and After Floating Measurement Operating Basics 2Ć18 ...
Page 111: ...Reference ...
Page 112: ......
Page 174: ...Pulse Mode Reference 3Ć62 ...
Page 176: ...Auxiliary Voltage Supply Reference 3Ć64 ...
Page 186: ...Adjusting the Graph Display Reference 3Ć74 ...
Page 190: ...Mathematical Operation Reference 3Ć78 ...
Page 210: ...Zoom Reference 3Ć98 ...
Page 217: ...Text Editing on the Graph 372 User Manual 3Ć105 Figure 3Ć46 ăClear Labels PopĆup Menu ...
Page 218: ...Text Editing on the Graph Reference 3Ć106 ...
Page 228: ...Hardcopy Reference 3Ć116 ...
Page 230: ...Initializing the 372 Reference 3Ć118 ...
Page 234: ...Time Stamp Reference 3Ć122 ...
Page 236: ...Adjusting CRT Brightness Reference 3Ć124 ...
Page 238: ...ID Information Reference 3Ć126 ...
Page 254: ...Floppy Disk System Reference 3Ć142 ...
Page 280: ...Sample Programs Reference 3Ć168 ...
Page 286: ...GPIB Reference 3Ć174 ...
Page 287: ...Appendices ...
Page 288: ......
Page 292: ...Appendix B Specifications Appendices AĆ4 ...
Page 312: ...Appendix B Specifications Appendices AĆ24 ...
Page 358: ...Appendix B Specifications Appendices AĆ70 ...