Data
Value Unit
Weight (without patient bed)
depending on configuration
640 - 690 kg
Length (without patient bed)
208 cm
Length (with patient bed)
210 cm
Width (without patient bed)
83 cm
Height
168 cm
11.2
Operating conditions
Data
Value Unit
Ambient temperature
18
–
24 °C
Relative humidity, maximum (non-condensing)
30
–
50 %
Air pressure
800
–
1060 hPa
Air purity
no dust; no
smoke; no evap‐
orating cleaning
agents; no
chemical or
alcoholic liquids
Floor loading, minimum
0.9 kg/cm
2
The room temperature has to be kept stable with a maximum change of 1 °C per
hour. Therefore active air conditioning is required.
The lighting of the operating room must comply with lighting standards for oper‐
ating rooms.
Data
Value Unit
Maximum operating time without shutdown of
the laser system
If operating con‐
ditions are
within the previ‐
ously specified
parameters, shut
down the laser
system if it will
not be used for
the next 24
hours.
Operating life
5 years
Ambient operating condi‐
tions
Operating time
TECHNICAL DATA
11
SKU: 70005780TPV * Version 6 * UM-100010770 GB
11 - 3
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 193 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...