Performing a therapeutic treatment
Materials:
n
Patient Interface 125 of the Patient Interface 125 Kit
(REF 90000115)
or
n
Patient Interface 125 of the Patient Interface 125 Kit
(REF 90000145)
n
Suction Clip 125 of the Patient Interface 125 Kit (REF
90000115)
or
n
Suction Clip 125 of the Patient Interface 125 Kit (REF
90000145)
The following chapter describes a penetrating keratoplasty procedure as an example
for performing a therapeutic treatment.
The performing of a lamellar keratoplasty procedure, a crosslinking procedure or an
ICRS tunnel procedure is similiar to the penetrating keratoplasty procedure, except the
differences in the respective procedures that are described in detail.
The surgeon must observe the treatment via the video live image in order to
detect potential problems.
1.
Perform steps 1 to 6 in
Ä
Chapter 9.5 ‘Performing a cataract treatment’
on page 9 - 30
.
2.
Verify that the ‘PI suction’ status indicator, the ‘Eye suction’ status indi‐
cator and the pressure status display are all green (Fig. 165).
Fig. 165: System parameters are verified
L
CAUTION
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 55
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 181 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...