4.
Click the
‘OK’
button to close the message ‘Power test successful.’ .
ð
The ‘Power test’ screen closes and the status indicator turns green
(Fig. 108).
9.3.6
Performing the depth test
Special tool:
n
Long adapter
n
Round plate holder
n
Support block
Materials:
n
Patient Interface 125 of the Patient Interface 125 Kit
(REF 90000115)
or
n
Patient Interface 125 of the Patient Interface 125 Kit
(REF 90000145)
You can use a PI 125 for up to five depth tests. To use a PI 125 more than once, choose
a different diameter in the
‘Diameter:’
field for each depth test.
1.
Click the
‘System test’
button (Fig. 27/5) on the ‘Main menu’ screen.
2.
Click the
‘Depth test’
button (Fig. 45/6) to open the ‘Depth test’ screen.
Fig. 109:
‘Depth test’
‘Preparation (1/3)’
screen
Fig. 108:
‘Power test’
successful
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 7
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 133 of 207