7.
Verify that the ‘PI suction’ status indicator, the ‘Eye suction’ status indi‐
cator and the pressure status display are all green.
Fig. 135: System parameters are verified
9.5.2
Performing a cataract treatment supported by a 2S-OCT
1.
To confirm, click the
‘Approve’
button (Fig. 135). The automatic identifica‐
tion of ocular structures is performed.
Fig. 136:
‘LRCS’
‘Treatment alignment - lens’
screen
ð
The yellow marking circle appears in the video camera image (Fig. 136).
The pupil margin is determined.
The yellow marking outlines appear in the linear scan images (Fig. 136).
The lens margins are determined.
The capsulotomy appears in the linear scan images and in the ring scan
image (Fig. 136).
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 31
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 157 of 207