To activate the vacuum for the suction clip (REF 90000145), press the
‘Vacuum’ footswitch once. Alternatively enable the ‘Eye suction’ (Fig. 130)
in the software.
Fig. 131:
‘Regular docking and treatment alignment’
screen:
‘Eye suction’
enabled (example)
ð
The suction reservoir and the suction port must be properly connected to
prevent suction loss. If a suction loss has occurred, the
‘Eye suction’
status
indicator turns gray. The procedure(s) cannot be performed.
14.
If a suction loss has occurred, check the connection between the suction
port and the suction reservoir. Adjust the connection, if necessary, and
repeat steps 10 to 13.
15.
To verify that the suction is actually engaged, gently lift the suction clip
(REF 90000145).
16.
Press the release button at the patient bed and wait 3 seconds for the lock
to disengage, swing the patient bed completely in (under the laser system)
and wait 3 seconds for the lock to engage.
9.4.4
Docking the patient to the spacer cone
Depending on the
headrest
being used, dock the patient to the spacer cone as
described in the following subchapters.
Fig. 130:
‘Eye suction’
disabled
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 22
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 148 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...