8.
Confirm with
‘Accept’
(Fig. 141).
If the anterior lens 0° position and the posterior lens 0° position are not in the
right position, click the ‘Edit’ button again. Mark the anterior lens 0° position
and the posterior lens 0° position to find the right positions.
9.
To mark the anterior lens 90° position and the posterior lens 90° position,
perform the steps as described in step 5 to 8 by using the 90° images.
The yellow marking outline representing the selected surface appears in both
the
‘Linear scan 0°’
and
‘Linear scan 90°’
OCT B scan images if the positions
have been marked in both images. If the positions have been marked in only
one image, they are indicated by the green marking dots; the yellow outline is
not displayed.
10.
Click the
‘Mark capsulotomy’
button in the ring scan OCT B scan image
(Fig. 137), if it is necessary to correct the capsulotomy.
Mark the highest point and the lowest point of the anterior capsule of the
lens, if necessary.
11.
To increase the rim height, click the
‘+’
button (Fig. 142).
To decrease the rim height, click the
‘-’
button (Fig. 142).
Fig. 142: Capsulotomy setting is performed
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 36
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 162 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...