4.
Shut down the laser system (
Ä
Chapter 6.2.3.1 ‘Switching the laser system on
and off’ on page 6 - 4
).
9.3
Performing the daily system test
Whenever you change the treatment mode from
‘LRCS/Therapeutics (80 kHz)’
to
‘Flap (160 kHz)’
or vice versa during a treatment day, you must
always
repeat the
following tests:
–
Select the treatment mode ‘LRCS/Therapeutics (80 kHz)’ for cataract/therapeutic
treatments or ‘Flap (160 kHz)’ for flap procedures (Fig. 27/4) in the
‘Main menu’
screen.
–
Perform the following tests:
–
‘Laser position test’
–
‘Depth test’
–
‘Power test’
9.3.1
Chiller submenu
1.
Click the
‘System test’
button (Fig. 27/5) on the ‘Main menu’ screen.
2.
Click the
‘Chiller’
button (Fig. 45/1) to open the ‘Chiller’ screen.
Fig. 100:
‘Chiller’
screen after chiller functionality test
ð
The ‘Chiller’ screen (Fig. 100) opens.
The following information are displayed:
n
‘Connection’ status
n
‘Power’ status
n
‘Coolant level’ status
If no further treatments will
be performed on that day
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 2
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 128 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...