6.2.6.1
Switching the laser ring light illumination on and off
1.
Switch on the laser ring light illumination by turning the knob as follows
(Fig. 12/1):
n
Position ‘2’ : 2 LEDs are illuminated (for less reflection while centering
the pupil)
n
Intermediate positions between ‘2’ and ‘OFF’
n
Intermediate positions between ‘OFF’ and ‘10’
n
Position ‘10’ : 10 LEDs are illuminated (for centering the pupil while
docking)
2.
Switch off the laser ring light illumination by turning the knob to the ‘OFF’
position.
6.2.7
Camera focus switch
The camera focus switch (Fig. 12/2) allows to shift the camera focus for corneal
and lens treatments.
6.2.7.1
Adjusting the camera focus
By turning the switch ‘CAM FOCUS’ (Fig. 12/2) clockwise the OP camera
focuses on the lens, and by turning the switch counterclockwise the OP
camera focuses on the cornea.
6.2.8
Spacer cone
Property damage due to an unprotected spacer cone!
The spacer cone is very delicate and may be damaged when it is not pro‐
tected.
– Always protect the spacer cone with its protection cover when the laser
system is not in use.
The spacer cone is the exit aperture of the laser beam. During treatment, the PI
125 and the patient are docked to the spacer cone via a vacuum.
NOTICE
HARDWARE
6
SKU: 70005780TPV * Version 6 * UM-100010770 GB
6 - 7
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 55 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...