The result of each system test is displayed in the
‘System test’
screen next to the rele‐
vant status indicator (Fig. 46).
Performing the
‘System test’
How to perform these system tests is described in the
Ä
Chapter 9.3 ‘Performing the
daily system test’ on page 9 - 2.
1.
Click the
‘System test’
button on the ‘Main menu’ screen (Fig. 27/5) to open
the ‘System test’ screen (Fig. 45).
Fig. 45:
‘System test’
screen
2.
Click the relevant button (Fig. 45) to open the dialog box for the associated
system test screen.
You can proceed as folllows:
n
Perform the ‘Laser position test’
n
Perform the startup tests
n
Verify the results of the automatic functionality test for the chiller.
Status indicator
Explanation
gray
System test has not been performed yet.
green
Test parameters are within the acceptable range. You can
proceed with the next test or procedure.
red
A system error has occured. You will not be able to start
with the procedure. Contact an Authorized Service
Technician.
Opening the ‘System test’
screen
Status indicators
Fig. 46:
‘System test’
status indicators
SOFTWARE
8
SKU: 70005780TPV * Version 6 * UM-100010770 GB
8 - 22
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 88 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...