8.
When the flap procedure is finished, the ‘Flap’ progress bar is highlighted
blue, the information footer shows the message ‘Treatment completed’
(Fig. 164), the acoustic signal stops and the green highlighted frame disap‐
pears in the ‘Treatment’ screen.
Fig. 164: Flap procedure is completed
9.
Click the
‘Continue’
button (Fig. 164) to open the ‘Treatment result’ screen
that shows the procedure results and the number of remaining licenses.
ð
L
CAUTION! Danger of trauma due to lowering the patient bed
without having released the vacuum of the suction clip!
When the vacuum is activated, lowering the patient bed leads to
an abrupt loss of vacuum and causes pain to the patient.
– Do not lower the patient bed while the vacuum is activated and
the patient is docked.
10.
Do not remove the suction clip while the vacuum is activated.
11.
Verify that the eye suction has been released after the procedure has been
executed.
ð
The ‘Eye suction’ status indicator of the suction clip turns gray
.
If a treatment is completed and the
‘Eye suction’
status indicator does not
turn gray, you must manually pull off the suction reservoir from the suction
port at the laser system.
12.
Disable the ‘PI suction’ in the software to deactivate the vacuum of the
spacer cone.
13.
Complete the flap procedure (
Ä
Chapter 9.4.6 ‘Completing a treatment’
on page 9 - 29
).
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 54
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 180 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...