13.
Press the
‘Start’
button to start the treatment (Fig. 143). When the green
highlighted frame appears in the ‘Treatment’ screen, finally press the
‘Procedure’ footswitch.
The
‘Procedure’
footswitch must be pressed from the start of the capsulotomy
procedure until the end of the lens fragmentation procedure! If you release the
‘Procedure’
footswitch, the treatment will pause. Press the footswitch again to
continue the treatment.
ð
An acoustic signal is emitted during the procedures.
The progress is displayed in the
‘Capsulotomy’
progress bar (Fig. 144).
L
WARNING! Risk of tissue bridges due to correction of vertical /
shear force movement!
The vertical force may change, or the applied shear force may
change during the procedure. If you attempt to correct such
changes, this may affect the alignment or the treatment result,
which may cause severe patient injuries.
– Do not correct changing vertical force or shear force during the
procedure.
When the capsulotomy procedure is finished, the lens fragmentation
procedure starts automatically. The progress is displayed in the ‘Lens
fragmentation’ progress bar (Fig. 144).
Fig. 144: Capsulotomy procedure and lens fragmentation procedure completed
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 38
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 164 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...