Parameter dialog box
Definition
‘Bottom bonus [µm]:’
Extension of the rim cut inside the cornea to
ensure that rim and bed cut overlap.
‘Arc length [°]:’
Arc length of rim cut, normally 360°.
‘Arc center position [°]:’
Central position of rim cut in case it is not 360°.
‘Posterior diameter [mm]:’
Diameter of the LKP bed resulting from
entered pachymetry, diameter and side cut
angle.
8.5.8
Parameters for crosslinking
Precondition:
n
A crosslinking procedure must be selected (Fig. 87).
For general information about how to view or specify parameter values refer to
Ä
Chapter 8.5.1 ‘Parameters for capsulotomy’ on page 8 - 36. Parameter-specific
information are described below.
Fig. 87: Crosslinking dialog boxes
‘Size’ parameter values
Parameter dialog box
Definition
‘Depth [µm]:’
Depth of bed and incisions.
‘Diameter [mm]:’
Diameter of bed cut.
‘Size’ dialog box
Fig. 88:
‘Size’
dialog box
SOFTWARE
8
SKU: 70005780TPV * Version 6 * UM-100010770 GB
8 - 53
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM