Parameter dialog box
Definition
‘Diameter [mm]:’
Diameter of the LKP at the anterior surface of
the cornea.
‘Pachymetry [µm]:’
Pachymetry of the thinnest point of the cornea
at the diameter of the LKP.
‘Depth ratio [%]:’
Ratio of cutting depth to pachymetry.
‘Resulting depth [µm]:’
Cutting depth resulting from pachymetry and
depth ratio.
‘Posterior diameter [mm]:’
Diameter of the LKP bed resulting from
entered pachymetry, diameter and side cut
angle.
Parameter dialog box
Definition
‘Energy [µJ]:’
Energy level of each single pulse when cutting
the bed.
‘Spot spacing [µm]:’
Distance between adjacent spots on a line
when cutting the bed.
‘Line spacing [µm]:’
Distance between spots of adjacent lines
when cutting the bed.
Parameter dialog box
Definition
‘Energy [µJ]:’
Energy level of each single pulse when cutting
the rim.
‘Spot spacing [µm]:’
Distance between adjacent spots on a line
when cutting the rim.
‘Line spacing [µm]:’
Distance between spots of adjacent lines
when cutting the rim.
‘Side cut angle [°]:’
Angle between the corneal surface tangent
and the rim cut. A side cut angle below 90°
leads to larger anterior than posterior diam‐
eter.
‘Top bonus [µm]:’
Extension of the rim cut at the anterior side of
the cornea to ensure exit of laser cut out of the
eye.
‘Size’ dialog box
Fig. 84:
‘Size’
dialog box
‘Bed’ dialog box
Fig. 85:
‘Bed’
dialog box
‘Rim’ dialog box
Fig. 86:
‘Rim’
dialog box
SOFTWARE
8
SKU: 70005780TPV * Version 6 * UM-100010770 GB
8 - 52
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 118 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...