8.1.5
Checking treatment licenses
Special tool:
n
User key
1.
Insert the user key in the ‘User Key’ port (Fig. 41/1) at the start-up panel of
the laser system.
2.
Click the
‘Licenses’
button on the ‘Main menu’ screen (Fig. 27/7).
ð
The ‘Remaining licenses’ dialog box (Fig. 42) shows the available/ena‐
bled procedures and the respective number of available licences.
3.
To close the ‘Licenses’ screen, click the
‘Main menu’
button.
8.1.6
Restarting the software after an emergency shutdown
Ensure that no patient or equipment are attached to the spacer cone!
1.
If required, undock the patient or the equipment and click
‘Confirm’
.
ð
The software will shut down. Wait for the forced shutdown to close all
programs.
2.
Press the ‘Reset’ button (Fig. 41/5) at the start-up panel.
3.
Wait for approximately 30 minutes until the reconnection is completed.
4.
Continue as described in
Ä
Chapter 8.1.1 ‘Starting and closing the software’
on page 8 - 1
.
Fig. 41: Start-up panel -
‘User Key’
port
Fig. 42:
‘Licenses’
screen /
‘Remaining
licenses’
dialog box
SOFTWARE
8
SKU: 70005780TPV * Version 6 * UM-100010770 GB
8 - 18
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 84 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...