‘Laser Stop’ button with integrated ‘Bed Stop’
Action
Results
To activate the ‘Laser Stop’
button:
n
Press the ‘Laser Stop’
button .
Any possible laser emission stops immediately by cut‐
ting off the laser interlock.
The patient bed cannot be moved upwards or down‐
wards.
The integrated illumination in the ‘Laser Stop’ button is
no longer illuminated.
The ‘Reset’ button is illuminated simultaneousely.
To unlock the ‘Laser Stop’
button:
n
Pull the ‘Laser Stop’
button towards you so
that a restart is possible.
n
Press the ‘Reset’ button
(Fig. 9/5).
After 2 seconds the ‘Reset’ button is no longer illumi‐
nated (normal operation).
The integrated illumination in the ‘Laser Stop’ button is
illuminated red (normal operation).
Depending on the
headrest
connected to the patient bed, please note the following
warning!
Headrest with external joysticks
Headrests with integrated joysticks
Red pushbutton at the headrest (at
the patient bed ‘LS Comfort’
Pushbutton at the headrest (at the
patient bed ‘Patient Support System
S60’ )
Patient beds with headrests
SAFETY
2
SKU: 70005780TPV * Version 6 * UM-100010770 GB
2 - 10
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 26 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...