10
Troubleshooting chart
Danger of injury due to improper repair!
Improperly executed work in conjunction with troubleshooting may lead to
serious injuries and significant property damage.
– Perform only troubleshooting tasks that local personnel can perform
according to the troubleshooting chart.
– If you are in doubt about how to perform the troubleshooting tasks that
are described in the troubleshooting chart, contact an Authorized
Service Technician.
– Troubleshooting tasks that are not described in the troubleshooting
chart must be performed only by Authorized Service Technicians.
Fault description
Cause
Remedy
Personnel
Laser system cannot be
started by turning the system
key to the ‘On’ position.
The main switch on the
bottom side of the laser
system is switched off
(position ‘0’ ).
n
Press the main switch to posi‐
tion ‘I’ .
n
Turn the system key to the
‘On’ position.
Assistant
User
Laser system is not ready for
operation, although power is
on.
‘Laser Stop’ button is acti‐
vated.
n
Ensure that the cause of the
laser stop has been removed
before unlocking the ‘Laser
Stop’ button.
Assistant
User
Authorized
Service Tech‐
nician
‘Reset’ button is illuminated
blue.
n
Press the blue ‘Reset’ button
until it is not illuminated any‐
more.
Assistant
User
Patient bed has not been ref‐
erenced.
n
Reference the patient bed.
Assistant
User
Laser system is not ready for
operation. ‘System test’
status indicator is illuminated
orange or gray.
Functionality tests have not
been performed successfully.
n
Contact an Authorized Service
Technician.
Authorized
Service Tech‐
nician
Startup tests have not been
performed successfully.
n
Contact an Authorized Service
Technician.
Authorized
Service Tech‐
nician
Not all startup tests have
been performed.
n
Perform the missing startup
tests.
Assistant
User
‘Laser position test’ failed.
The laser source has not
warmed up / stabilized yet.
n
If the ‘Laser position test’ fails
repeatedly, even after 20
minutes warm-up, contact an
Authorized Service Techni‐
cian.
Authorized
Service Tech‐
nician
L
WARNING
Laser system in general
TROUBLESHOOTING CHART
10
SKU: 70005780TPV * Version 6 * UM-100010770 GB
10 - 1
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 187 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...