Pos.
Button/Field
Description / Action
10
‘Linear scan 0°’ OCT B
scan image
Shows the real-time OCT linear scan image in 0° position of the patient's
eye.
The flashing camera symbol represents the periodical updating of the OCT
B scan image.
Click the
‘Edit’
button to manually mark the anterior lens 0° position and
posterior lens 0° position.
11
‘Linear scan 90°’ OCT B
scan image
Shows the real-time OCT linear scan image in 90° position of the patient's
eye.
The flashing camera symbol represents the periodical updating of the OCT
B scan image.
Click the
‘Edit’
button to manually mark the anterior lens 90° position and
posterior lens 90° position.
12
Ring scan image
Provides depth information of structures in the eye scanned in a circular
manner at the location of the planned capsulotomy.
Click the
‘Mark capsulotomy’
button to mark the highest and lowest point
of the anterior lens surface in the ring scan to ensure a complete capsu‐
lotomy cut.
13
‘Confirm’
button
Click the
‘Confirm’
button to accept all workflow settings.
14
‘-’
/
‘+’
buttons
To increase the rim height of the capsulotomy, click the button.
To decrease the rim height of the capsulotomy, click the button.
15
‘Center to lens apex’
button
To center the capsulotomy and the lens fragmentation procedures to the
lens apex, click the button.
16
‘3-click-centration’
button
To mark the pupil margin, click the
‘3-click-centration’
button.
17
‘Cyclotorsion’ buttons
To rotate the arcuate incisions and the corneal incisions clockwise, click the
button.
To rotate the arcuate incisions and the corneal incisions counterclockwise,
click the button.
18
‘AI @ 225°’ ‘AI @ 45°’
OCT B scan images
These OCT B scan images show the position of the arcuate incisions.
To adjust the cutting depth of each arcuate incision, click the relevant
‘Edit’
button.
19
‘Incision @ 90°’ /
‘Incision @ 180°’ /
‘Incision @ 0°’ OCT B
scan images
These OCT B scan images show the cross sections of the cornea where the
corneal incisions are planned.
20
‘Continue with not
performed procedures’
button
To proceed with skipped procedures, click the
‘Continue with not performed
procedures’
button.
SOFTWARE
8
SKU: 70005780TPV * Version 6 * UM-100010770 GB
8 - 14
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 80 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...