Parameter
Definition
þ
‘Position of incision 1 [°]:’
Perform incision 1.
‘Position of incision 1 [°]:’
Position incision 1.
þ
‘Position of incision 2 [°]:’
Perform incision 2.
‘Position of incision 2 [°]:’
Position incision 2.
‘Length of incision 1 [µm]:’
Length of incision 1.
‘Length of incision 2 [µm]:’
Length of incision 2.
‘Energy [µJ]:’
Energy level of each single pulse when cutting
the incisions.
‘Spot spacing [µm]:’
Distance between adjacent spots on a line
when cutting the incisions.
‘Line spacing [µm]:’
Distance between spots of adjacent lines
when cutting the incisions.
‘Top bonus [µm]:’
Extension of the incision cut at the anterior
side of the cornea to ensure exit of laser cut
out of the eye.
‘Bottom bonus [µm]:’
Extension of the incision cut inside the cornea
to ensure that incision and tunnel cut overlap.
Parameters that remain unchanged are not listed in the table below.
Fig. 94: ICRS dialog boxes - single tunnel segment
Fig. 93:
‘Incisions’
dialog box - full
tunnel
‘Tunnel’ / ‘Incisions’
dialog boxes
SOFTWARE
8
SKU: 70005780TPV * Version 6 * UM-100010770 GB
8 - 56
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 122 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...