8.5.4
Parameters for corneal incisions
Precondition:
n
A corneal incisions procedure must be selected (Fig. 70).
Fig. 70:
‘Corneal incisions’
parameters dialog box
For general information about how to view or specify parameter values refer to
Ä
Chapter 8.5.1 ‘Parameters for capsulotomy’ on page 8 - 36. Parameter-specific
information are described below.
The dialog box shows the automatically updated B scan images of the primary cor‐
neal incision (
‘Primary’
), the secondary incision 1 (
‘Side port 1’
)and/or the secon‐
dary incision 2 (
‘Side port 2’
) (Fig. 71).
Fig. 71:
‘Corneal incisions’
parameters dialog box
Parameters
Parameter dialog box
Definition
‘Primary’
Perform a primary corneal incision.
‘Outer width [mm]:’ ( ‘Primary’ )
Adjust the outer width of the primary incision according to the size
of the selected IOL.
‘Inner width [mm]:’ ( ‘Primary’ )
Adjust the inner width of the primary incision according to the size
of the selected IOL.
‘Axis [°]:’ ( ‘Primary’ )
Select the position of the primary incision. This value may be
changed according to the cyclorotation of the eye after the patient
has been docked.
‘Corneal incisions’ dialog
box
SOFTWARE
8
SKU: 70005780TPV * Version 6 * UM-100010770 GB
8 - 43
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 109 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...