Fault description
Cause
Remedy
Personnel
‘Depth test’ failed.
The depth test has failed
after selecting another diam‐
eter in the ‘Diameter:’ drop‐
down field.
n
Try to perform the depth test
again.
n
If the ‘Depth test’ fails repeat‐
edly, contact an Authorized
Service Technician.
Authorized
Service Tech‐
nician
Fault
description
Cause
Remedy
Personnel
Laser system
has been
aborted
during a
treatment.
Laser system has
been aborted, for
instance, by
pressing the
‘Stop’
button in
the GUI.
n
If a system abort occurs or has been performed during a
treatment, make sure the eye suction is automatically
switched off. Otherwise switch it off.
Lower the patient bed and swing it in the patient entry posi‐
tion.
Refer to the user manual "Laser Patient Bed LS Comfort" or
"Patient Support System PSS S60" by AKRUS GmbH & Co KG.
n
Ask the patient to leave the patient bed very carefully.
n
Solve the problem, if possible.
n
Otherwise call an Authorized Service Technician, if necessary.
User
Authorized
Service
Technician
Emergency
shutdown in
the laser
system.
Emergency shut‐
down has been
initiated auto‐
matically by fail‐
ures.
n
If an emergency shutdown occurs during a treatment, make
sure the eye suction is automatically switched off. Otherwise
switch it off.
Lower the patient bed and swing it in the patient entry posi‐
tion.
Refer to the user manual "Laser Patient Bed LS Comfort" or
"Patient Support System PSS S60" by AKRUS GmbH & Co KG.
n
Ask the patient to leave the patient bed very carefully.
n
Call an Authorized Service Technician.
Authorized
Service
Technician
Fault
description
Cause
Remedy
Personnel
Laser system
has been
aborted
during a
treatment.
Laser system has
been aborted,
for instance, by
pressing the
‘Stop’
button in
the GUI.
n
If a system abort occurs or has been performed during a
treatment, make sure that the eye suction is automatically
switched off. Otherwise switch it off.
Lower the patient bed and swing it in the patient entry posi‐
tion.
Refer to the user manual "Laser Patient Bed LS Comfort" by
AKRUS GmbH & Co KG.
n
Unfasten the VELCRO
â
strapping, if used.
User
Authorized
Service
Technician
For headrest with inte‐
grated joystick only:
For headrest with external
joysticks only:
TROUBLESHOOTING CHART
10
SKU: 70005780TPV * Version 6 * UM-100010770 GB
10 - 2
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 188 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...