3.
Enable the ‘PI suction’ in the software to activate the vacuum of the spacer
cone.
Fig. 117:
‘Regular docking and treatment alignment’
screen:
‘PI suction’
enabled (example)
ð
The PI 125 (REF 90000115) is sucked to the spacer cone. The green ‘PI
suction’ status indicator (Fig. 117) indicates that the vacuum is in the
valid range.
The PI 125 (REF 90000115) must be properly connected to the spacer cone
to prevent suction loss.
L
WARNING! Danger of injury due to contamination of the PI 125
(REF 90000115)!
If the PI 125 (REF 90000115) is contaminated during the docking
process, infection of the patient's eye may occur.
– Do not open the PI 125 Kit (REF 90000115) packaging until you
are ready to use it.
Fig. 116:
‘PI suction’
disabled
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 15
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 141 of 207