4.
Place the suction clip centrally on the patient's eye.
ð
The suction clip is positioned correctly.
5.
Repeat the steps in
Ä
Chapter 9.4.4 ‘Docking the patient to the spacer cone’
on page 9 - 22
.
9.4.6
Completing a treatment
Depending on the
headrest
being used, complete the treatment as described as fol‐
lows:
1.
Lower the patient bed very carefully.
2.
Unfasten the Velcro strapping, if used.
3.
Swing the bed out completely so that the patient can easily get up.
4.
Ask the patient to leave the patient bed.
5.
Remove the PI 125 from the spacer cone carefully by clicking the blue
button next to the ‘PI suction’ status indicator.
ð
The ‘PI suction’ status indicator
turns gray.
6.
Remove the suction reservoir from the suction port.
7.
L
WARNING! Infection hazard due to reuse of the PI 125 and suction
clip!
The PI 125 and the suction clip are single-use devices. They must not
be refurbished or reused. Neglecting this rule may result in infections
and severe patient injuries.
– Never reuse or refurbish the PI 125 or the suction clip!
n
Dispose of the PI 125 and the suction clip after use.
n
Use the PI 125 and the suction clip exclusively on one eye.
For using a headrest with
external joysticks only:
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 29
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 155 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...