4.
Click the
‘OK’
button to close the message ‘Laser position test successful.’ .
ð
The ‘Laser position test ’ screen closes and the status indicator turns
green (Fig. 102).
9.3.3
Performing the vacuum test
Special tool:
n
Suction reservoir
If the
‘Vacuum test’
has been performed successfully, it is valid for
24 hours
!
1.
Click the
‘System test’
button (Fig. 27/5) on the ‘Main menu’ screen.
2.
Click the
‘Vacuum test’
button (Fig. 45/2) to open the ‘Vacuum test’ screen.
Fig. 103:
‘Vacuum test’
‘Preparation’
screen
3.
Follow the instructions in the ‘Vacuum test’ screen.
Fig. 102:
‘Laser position test’
suc‐
cessful
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 4
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 130 of 207