Parameter
Definition
Perform a single tunnel segment.
‘Tunnel width [µm]:’
( ‘Tunnel 1’ )
Width of the tunnel segment.
‘Tunnel length [°]:’ ( ‘Tunnel 1’ ) Length of the tunnel segment.
‘Position [°]:’ ( ‘Tunnel 1’ )
Central position of the tunnel segment.
Parameter
Definition
þ
‘Position of incision 1 [°]:’
Perform incision 1.
‘Length of incision 1 [µm]:’
Length of incision 1.
‘Position of incision 1 [°]:’
Position of incision 1.
o
‘Position of incision 2 [°]:’
Perform incision 2 (deactivated by default).
‘Length of incision 2[µm]:’
Length of incision 2 (deactivated by default).
‘Position of incision 2 [°]:’
Position of incision 2 (deactivated by default).
Parameters that remain unchanged are not listed in the table below.
Fig. 97: ICRS dialog boxes - two-tunnel segments
Fig. 95:
‘Tunnel’
dialog box - single
tunnel segment
Fig. 96:
‘Incisions’
dialog box - single
tunnel segment
‘Tunnel’ / ‘Incisions’
dialog boxes
SOFTWARE
8
SKU: 70005780TPV * Version 6 * UM-100010770 GB
8 - 57
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 123 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...