WARNING! Risk of tissue bridges due to movement when laser is firing!
Tissue bridges may occur if vibrations from movement occur when the laser
is firing.
– Do not walk around while the procedure is performed.
– Do not touch the patient bed or the patient's head while the procedure is
performed.
– Do not touch the suction clip.
8.
When you have verified the parameter values, press the
‘Start’
button
(Fig. 167). When the green highlighted frame appears in the ‘Treatment’
screen, finally press the ‘Procedure’ footswitch. Remember to keep the
footswitch pressed during the whole procedure.
The
‘Procedure’
footswitch must be pressed from the start of the procedure
until the end of the procedure! If you release the
‘Procedure’
footswitch, the
treatment will pause. Press the footswitch again to continue the treatment.
ð
An acoustic signal is emitted during the procedure.
The progress is displayed in the
‘Penetrating keratoplasty’
progress bar
(Fig. 168).
L
WARNING! Risk of tissue bridges due to correction of vertical /
shear force movement!
Vertical force may change, or the applied shear force may change
while the procedure is being performed. If you attempt to correct
such changes, this may affect the alignment or the treatment
result, which may cause severe patient injury.
– Do not correct changing vertical force or shear force during the
treatment.
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 58
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 184 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...