8
Software
8.1
Working with the software
Navigation and editing of elements within the graphical user interface are identical to
a Microsoft
â
Windows operating system.
The laser system software can be operated from either of the two locations:
n
Assistant workstation
n
Surgeon control screen
The laser system software can be operated and managed by different personnel
(
Ä
Chapter 2.6 ‘Personnel requirements’ on page 2 - 4).
Fig. 20: Error message (example)
Problems are indicated by error messages.
8.1.1
Starting and closing the software
Danger of property damage due to aborted warm-up of the laser system!
The laser system must only be used after the warm-up has been completed.
If you do not allow the laser system to warm up properly, the entire laser
system may be permanently damaged.
– Do not abort warm-up.
– For initial start-up, wait at least 20 minutes after having turned the
system key to the ‘On’ position. Only after waiting that 20 minutes,
press the blue ‘Reset’ button. After having pressed the ‘Reset’ button,
wait for at least 20 minutes (automatic countdown).
Precondition:
n
The laser system and the monitors are switched on (
Ä
Chapter 6.2.3.1
‘Switching the laser system on and off’ on page 6 - 4
).
Operating the software
Error messages
Starting the software
NOTICE
SOFTWARE
8
SKU: 70005780TPV * Version 6 * UM-100010770 GB
8 - 1
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 67 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...