L
WARNING! Infection hazard due to loss of sterility!
The following task must be performed by the sterile user only to ensure the
sterility of the suction clip (REF 90000115).
8.
If the suction clip (REF 90000115) is not open, open it by pulling the snap-fit
(Fig. 120/1) backwards.
9.
Place the suction clip (REF 90000115) centrally onto the patient's eye.
ð
The suction clip (REF 90000115) is positioned correctly.
10.
To activate the vacuum for the suction clip (REF 90000115), press the
‘Vacuum’ footswitch once. Alternatively enable the ‘Eye suction’ (Fig. 121)
in the software.
Fig. 122:
‘Regular docking and treatment alignment’
screen:
‘Eye suction’
enabled (example)
ð
The suction reservoir and the suction port must be properly connected to
prevent suction loss. If a suction loss has occurred, the
‘Eye suction’
status
indicator turns gray. The procedure(s) cannot be performed.
11.
If a suction loss has occurred, check the connection between the suction
port and the suction reservoir. Adjust the connection, if necessary, and
repeat steps 5 to 10.
12.
To verify that the suction is actually engaged, gently lift the suction clip
(REF 90000115).
13.
Press the release button at the patient bed and wait 3 seconds for the lock
to disengage, swing the patient bed completely in (under the laser system)
and wait 3 seconds for the lock to engage.
Fig. 120: Suction clip with snap-fit
(REF 90000115)
Fig. 121:
‘Eye suction’
disabled
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 17
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 143 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...