Parameter dialog box
Definition
‘Energy [µJ]:’
Energy level of each single pulse when cutting
the bed.
‘Spot spacing [µm]:’
Distance between adjacent spots on a line
when cutting the bed.
‘Line spacing [µm]:’
Distance between spots of adjacent lines
when cutting the bed.
Parameter dialog box
Definition
0
Perform no incisions.
1
Perform one incision.
2
Perform two incisions.
‘Energy [µJ]:’
Energy level of each single pulse when cutting
the incisions.
‘Side cut angle [°]:’
Angle between the corneal surface tangent
and the incision cut. A side cut angle below
90° leads to larger anterior than posterior
diameter.
‘Spot spacing [µm]:’
Distance between adjacent spots on a line
when cutting the incisions.
‘Line spacing [µm]:’
Distance between spots of adjacent lines
when cutting the incisions.
‘Top bonus [µm]:’
Extension of the incision cut at the anterior
side of the cornea to ensure exit of laser cut
out of the eye.
‘Bottom bonus [µm]:’
Extension of the incision cut inside the cornea
to ensure that incision and bed cut overlap.
‘Arc length [mm]:’
Arc length of incision(s).
‘Position 1 [°]:’
Central position of incision 1.
‘Position 2 [°]:’
Central position of incision 2.
8.5.9
Parameters for ICRS tunnel
Precondition:
n
An intracorneal ring segments tunnel procedure must be selected (Fig. 91).
‘Bed’ dialog box
Fig. 89:
‘Bed’
dialog box
‘Incisions’ dialog box
Fig. 90:
‘Incisions’
dialog box
SOFTWARE
8
SKU: 70005780TPV * Version 6 * UM-100010770 GB
8 - 54
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 120 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...