10.
Perform steps 10 to 29 in
Ä
Chapter 9.5.2 ‘Performing a cataract treatment
supported by a 2S-OCT’ on page 9 - 31
.
9.6
Performing a corneal treatment
9.6.1
Performing a flap procedure
Materials:
n
Patient Interface 125 of the Patient Interface 125 Kit
(REF 90000115)
or
n
Patient Interface 125 of the Patient Interface 125 Kit
(REF 90000145)
n
Suction Clip 125 of the Patient Interface 125 Kit (REF
90000115)
or
n
Suction Clip 125 of the Patient Interface 125 Kit (REF
90000145)
The surgeon must observe the treatment via the video live image in order to
recognize possible problems.
1.
Perform steps 1 to 6 in
Ä
Chapter 9.5 ‘Performing a cataract treatment’
on page 9 - 30
.
2.
Verify that the ‘PI suction’ status indicator, the ‘Eye suction’ status indi‐
cator and the pressure status display are all green (Fig. 161).
Fig. 161:
‘Flap’
‘Regular docking and treatment alignment’
screen
ð
The
‘Start’
button becomes available (Fig. 161).
L
CAUTION
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 51
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 177 of 207