The illumination unit is positioned inside the laser system. It will be automatically
switched on when the laser system is started.
The footswitch (Fig. 18) adjusts the brightness of the surgical microscope. The left
pedal of the footswitch is used to darken the picture or to switch off the surgical
microscope; the right pedal is used to increase the brightness. The brightness
increment is more finely graduated and no longer linear.
The cable at the footswitch is connected to the surgical illumination unit.
If the laser system is running over night, remember to switch off the surgical micro‐
scope illumination by pressing the left footswitch (Fig. 18)!
6.2.10
Footswitches
‘Procedure’ footswitch
‘Vacuum’ footswitch
The ‘Procedure’ footswitch has a red
label and starts and controls the proce‐
dures.
The ‘Vacuum’ footswitch has a blue
label and activates/deactivates the
vacuum of the suction clip.
It is impossible to deactivate the vacuum of the suction clip with the
‘Vacuum’
foot‐
switch during the treatment in order to avoid suction loss, which would lead to proce‐
dure abort.
Aborting the treatment by pressing the ‘Stop’ button in the GUI will automatically
deactivate the vacuum.
For surgical microscope illu‐
minations that are inte‐
grated in the laser system:
Fig. 18: Footswitch for adjusting the sur‐
gical microscope illumination
HARDWARE
6
SKU: 70005780TPV * Version 6 * UM-100010770 GB
6 - 11
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 59 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...