WARNING! Infection hazard due to loss of sterility!
The following tasks must be performed by the sterile user only to ensure the
sterility of the suction clip (REF 90000145).
Fig. 127: Suction clip (REF 90000145)
10.
Without touching the suction clip blister pack (Fig. 123/2), take the suction
reservoir (Fig. 127/1) of the suction clip (REF 90000145) and connect it to
the suction port (Fig. 128).
Remove the suction clip (REF 90000145) by using the handle (Fig. 127/4).
Inspect it for any visible damage or contamination.
n
If the suction clip (REF 90000145) is damaged in any way, dispose of the
suction clip (REF 90000145) and use a different one.
L
WARNING! Infection hazard due to loss of sterility!
The following task must be performed by the sterile user only to ensure the
sterility of the suction clip (REF 90000145).
11.
If the suction clip (REF 90000145) is not open, open it by pulling the snap-fit
(Fig. 129) backwards.
12.
Place the suction clip (REF 90000145) centrally onto the patient's eye.
ð
The suction clip (REF 90000145) is positioned correctly.
Fig. 128: Suction reservoir connected
Fig. 129: How to open the suction clip
with snap-fit (REF 90000145)
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 21
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 147 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...