Fault
description
Cause
Remedy
Personnel
n
Ask the patient to leave the patient bed very carefully.
n
Solve the problem, if possible.
n
Otherwise call an Authorized Service Technician, if necessary
Emergency
shutdown in
the laser
system.
Emergency shut‐
down has been
initiated auto‐
matically by fail‐
ures.
n
If an emergency shutdown occurs during a treatment, make
sure that the eye suction is automatically switched off. Other‐
wise switch it off.
Lower the patient bed and swing it in the patient entry posi‐
tion.
Refer to the user manual "Laser Patient Bed LS Comfort" by
AKRUS GmbH & Co KG.
n
Unfasten the VELCRO
â
strapping, if used.
n
Ask the patient to leave the patient bed very carefully.
n
Call an Authorized Service Technician.
Authorized
Service
Technician
Fault description
Cause
Remedy
Personnel
‘Supply temperature’ and/or
‘Current flow’ are not in the
specified range.
Room temperature and/
or current flow are out‐
side of the specified
range.
n
Shut down the laser system and
try to restart it.
n
If the problem persists, contact
an Authorized Service Techni‐
cian.
Assistant
User
Authorized
Service Techni‐
cian
Fault description
Cause
Remedy
Personnel
LED bar display flashes at regular
intervals.
Error in the vacuum
pump unit.
n
Contact an Authorized
Service Technician.
Authorized Service
Technician
Fault description
Cause
Remedy
Personnel
Monitors are blank although power
is on.
Monitors have been switched off
separately.
n
Switch on the
monitors.
Assistant
User
Power cables / PC cables are not
plugged in.
n
Plug in the cables.
Assistant
User
Chiller
Suction status controller
Assistant workstation,
surgeon control screen
TROUBLESHOOTING CHART
10
SKU: 70005780TPV * Version 6 * UM-100010770 GB
10 - 3
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 189 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...