8.
Click the button next to the OCT B scan image (Fig. 158). To mark the
posterior lens 90° position, mark the posterior capsule of the lens with three
points.
Fig. 159: Posterior lens 90° positioning is performed
ð
The green marking line representing the selected surface, appears in
the OCT B scan image. The frame around the button is highlighted
green (Fig. 159).
If the posterior lens 90° is not in the right position, click once again on the
button. Mark the posterior lens 90° position as described to find the right
position.
9.
To activate the capsulotomy setting, click the button next to the OCT B
scan image (Fig. 159).
Fig. 160: Capsulotomy in the ring scan OCT B scan image
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 50
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 176 of 207