9.4.4.1
For capsulotomy and lens fragmentation procedures using a
headrest with integrated joystick
1.
Move the patient bed using the joystick (Fig. 132/1 or Fig. 133/1) of the
patient bed control panel at the headrest until the suction clip and the PI
125 are aligned.
2.
Instill a maximum of six drops of balanced salt solution (BSS) in the suction
clip.
3.
Slowly move the patient bed upward until the suction clip almost touches
the PI 125.
4.
Be aware that the reflection of the laser ring light is visible in the video camera
image.
Ensure that the laser ring light illumination switch is adjusted to position
‘10’
(Fig. 12/1). Align the reflection of the laser ring light on the cornea with
the laser ring light and gently move the patient bed upward.
5.
Slowly move the patient bed upward until the central cornea gently
touches the PI 125. Ensure that there are no wrinkles visible in the OCT
image.
L
CAUTION!
Ensure that the pressure status display does not reach the upper red
pressure display area! Otherwise there is a higher risk to have wrinkles
in the posterior cornea.
ð
The shear force display shows the lateral pressure that is exerted on the
eye. The optimal position is achieved when the LEDs are illuminated
green.
Fig. 132: Patient bed control panel at
the headrest (
‘LS Comfort’
)
Fig. 133: Patient bed control panel at
the headrest (
‘Patient Support System
S60’
)
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 23
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 149 of 207
Summary of Contents for Victus
Page 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Page 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Page 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Page 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Page 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...