CO
2
Monitoring
17-24
NOTE
z
Do not use the accessory if the packaging or the internal accessory is
damaged. Return it to the manufacturer.
z
“CO2 START UP” and “CO2 WARM UP” displayed on the screen indicate
that the sensor is starting up and warming up. During the warming up,
the module can measure CO
2
, but the measurement is not standard.
After the above information disappears, the standard measurement can
be performed.
17.4.3 CO
2
Setup Menu
Selecting the CO
2
label in the parameter window opens the following menu.
Figure 17-13 CO
2
Setup Menu
You can perform the following settings in this menu.
ALM
Alarm on/off status
ON: When a CO
2
alarm occurs, the monitor gives alarm
indications and stores the alarm;
OFF: When a CO
2
alarm occurs, the monitor neither gives
alarm indications nor stores the alarm;
When OFF is selected, the
icon is displayed on the
right of the CO
2
label.